Spec
|
HI 991300
|
HI 991301
|
Range:
|
|
|
pH
|
0.00 to 14.00 pH |
0.00 to 14.00 pH |
EC |
0 to 3999 uS/cm |
0.00 to 19.99 mS/cm |
TDS |
0 to 2000 ppm |
0.00 to 10.00 ppt |
degC / degF
|
0.0 to 60.0 degC (32 to 140degF) |
0.0 to 60.0 degC
(32 to 140degF) |
|
|
|
Resolution:
|
|
|
pH
|
0.01 pH |
0.01 pH |
EC |
1 uS/cm |
1 mS/cm |
TDS |
1 ppm |
1 ppt |
degC / degF
|
0.1 degC/degF
|
0.1 degC/degF |
|
|
|
Accuracy (@ 20 degC):
|
|
|
pH
|
+/-0.01 pH |
+/-0.01 pH |
EC/TDS |
2% Full Scale |
2% Full Scale |
degC / degF |
+/-0.5 degC (+/-1.0F) |
+/-0.5 degC (+/-1.0F)
|
|
|
|
Typical EMC Deviation:
|
|
|
pH
|
+/-0.3pH
|
+/-0.3 pH
|
EC/TDS
|
2% Full Scale |
2% Full Scale |
degC / degF |
+/-1 degC (+/-1.0 degF) |
+/-1 degC (+/-1.0 degF)
|
|
EC/TDS Calibration
|
1 point Automatic
at 1382 ppm (conv.=0.5);
1500 ppm (conv.=0.7);
1413 uS/cm (others) |
1 point Automatic
at 6.44 ppt (conv.=0.5);
9.02 ppt (conv.=0.7);
12880 uS/cm (others) |
|
|
pH Calibration
|
Automatic 1 or 2 points with
2 sets of memorized buffers
(4.01/7.01/10.01 or 4.01/6.86/9.18) |
|
|
EC/TDS Ratios |
Selectable between 0.45,
0.50 (default), 0.55, 0.60, 0.65, 0.68, 0.70, 0.75, 1.00 ppm=1 uS |
|
|
Temperature
Compensation
|
pH: Automatic from 0 to
60 degC
EC/TDS: Automatic from 0 to 60 degC with selectable B of
0.0, 1.8, 1.9 (default), 2.0, 2.1, 2.2, 2.3, 2.4% per degC |
|
|
Probe
|
HI 1288 pH/EC/TDS with
temp sensor, DIN connector and 1 m (3.3’) cable (included) |
|
|
Battery Life/Type |
500 hours of continuous use
/ 4 x 1.5V AAA |
|
|
Environment |
0 to 50 degC (32 to 122 degF);
RH 100% |
|
|
Dimensions |
143 x 80 x 38 mm (5.6 x 3.1
x 1.5’) |
|
|
Weight |
320g (11.3 oz) |